How to Set the Timer Controller of a Sand and Dust Test Chamber

Source:LINPIN Time:2025-05-06 Category:Industry News

How are the parameters of the sand and dust test chamber timer controller set? Here’s a guide: Today, we’ll share the detailed steps for operating the timer controller.

Before operating the timer controller of the sand and dust test chamber, ensure that the power is connected, as these operations must be performed with the power on. Initially, the controller display will show the main menu: Run, Function Block, Clear Program, etc.
sand and dust test chamber

Access the Function Block

Press the Up/Down Arrow keys to move the cursor to Function Block, then press OK (Confirm).
The screen will display “T1” (Test Time) in the lower right corner, with the cursor blinking on “T1”.
Set the Test Time (T1)

Press the Left/Right Arrow keys to move the blinking cursor to the time setting position.
Observe the time unit displayed in the upper left corner of the screen.
Press SEL, then use the Up/Down Arrow keys to adjust the desired time value.
Press OK to confirm, and the cursor will return to “T1”.
Set Blowing Dust Time (T2)

Press the Up Arrow and SEL keys simultaneously—the screen will display “T2” (Blowing Dust Time, recommended: 5 minutes).
Follow the same method as T1 to set the time.
Set Blowing Dust Stop Time (T3)

Press the Up Arrow—the screen will display “T3” (Stop Blowing Dust Time, recommended: 2 minutes).
Set it using the same method as T1.
Set Vibration and Power Times (T4-T7)

T4: Vibration Time (recommended: 0.5 seconds).
T5: Stop Vibration Time (recommended: 2 minutes).
T6: Power-On Time (set according to relevant standards).
T7: Power-Off Time (also set according to standards).
Complete Setup and Start the Test

After setting all 7 time parameters, press ESC (Exit) to return to the main menu.
Use the Up Arrow to move the cursor to “Run”, then press OK—”Run” will change to “Stop”, and the internal power will activate.
The temperature controller will display readings. Press ESC again to return to the control screen.
Begin the Test

Turn on the “Blowing Dust” and “Vibration” switches—the device will start working.
To begin the test, turn on the “Sample” switch.
That’s all for the operation steps of the sand and dust test chamber timer controller. By reading this guide, you should now have a clear understanding of the process.

News Recommendation
Temperature control devices in salt spray test chambers regulate temperature variations in the working environment. These devices utilize physical changes within their mechanisms to trigger specific responses, such as opening or closing circuits, thereby achieving automatic temperature control.
A high-temperature test chamber simulates extreme heat environments to evaluate the performance of test specimens. But how does prolonged exposure to high temperatures affect these specimens? This article explains the potential impacts of high temperatures on various materials.
Frost formation in a thermal shock test chamber is a sublimation phenomenon where moisture inside the chamber condenses under low-temperature conditions. But why does this happen? Is it simply because the chamber temperature is too low?
A HAST Accelerated Aging Test Chamber (Highly Accelerated Stress Test Chamber) is a laboratory device used to simulate an accelerated aging environment.
Salt-spray corrosion testing is the primary accelerated method for validating the corrosion resistance of materials and their protective coatings. Whether a chamber can continuously generate neutral (NSS), acetic-acid (AASS) or copper-accelerated acetic-acid (CASS) salt fog for 48 h–1 000 h determines the repeatability and reproducibility of the test.
Product Recommendation
Telegram WhatsApp Facebook VK LinkedIn