Expert Tips: How to Deal with Nozzle Clogging in a Salt Spray Test Chamber

Source:LINPIN Time:2025-05-20 Category:Industry News

If your salt spray test chamber experiences spray failure, the issue may be a clogged nozzle. So, how should you handle nozzle clogging? Here’s a clever trick from our experts—a simple guide on how to resolve nozzle blockages in testing equipment.

As the saying goes, “The right remedy works best.” To quickly fix a clogged nozzle in a salt spray test chamber, you first need to identify the cause of the blockage. Possible reasons include:

Long-term disuse and lack of cleaning, leading to crystallization or chemical deposits that form blockages in the nozzle or its channels (a coupling effect where two or more substances combine).
Frequent use with a damaged filtration system, causing fluid to create an abrasive “sandblasting” effect that damages the nozzle.

salt spray test chamber
Physical impact, such as cracks or chips in the nozzle due to accidental collisions.
To diagnose the issue, check the pressure gauge. If the pressure is normal, the problem likely lies in the nozzle or tubing. Observe whether the saline solution flows properly to determine whether the blockage is in the nozzle or the pipeline.

If the nozzle is clogged with debris:

Use an air gun to blow out the blockage from the nozzle’s tip (never blow from the silicone tube side).
If there is no debris but poor airflow:

Turn on the spray switch and check if air flows through the curved glass tube of the nozzle. If not, adjust the nozzle’s position.
If the nozzle is damaged:

Replace it immediately to ensure proper operation.
Preventive Measures to Avoid Nozzle Clogging
To minimize nozzle blockages:

Use filtered or softened water (such as distilled or deionized water) in the test chamber to reduce crystallization.
Avoid overly small nozzle apertures, as they are more prone to clogging.
Install an oil-water separator to prevent moisture and oil from the compressed air from contaminating the nozzle.
Routine maintenance and careful handling can prevent accidental collisions that may damage the nozzle.
By following these steps, you can effectively prevent and resolve nozzle clogging, ensuring smooth and accurate salt spray testing. Stay tuned for more maintenance tips on testing equipment!

News Recommendation
If a salt spray test chamber is used for a long time without cleaning, dust will inevitably accumulate both inside and outside the equipment.
The salt fog chamber is the core instrument for evaluating the corrosion resistance of materials and protective coatings. Its measurement repeatability and reproducibility are directly affected by installation quality.
During the past twelve months, several third-party testing bodies and OEM laboratories reported the same problem: when running dust tests in accordance with GB/T 4208, IEC 60529 or MIL-STD-810, the batch cycle has stretched from 48 h to 72 h, even 96 h, while energy consumption has risen by more than 30 %.
Today, the editor of Linpin Instruments will guide us in troubleshooting the alarm of the high - low temperature test chamber.
Chongqing Chuanyi Microcircuit Co., Ltd. (referred to as Chuanyi Micro) was established in 2002 and is a state-owned enterprise under the China Four United Group.
Product Recommendation
Telegram WhatsApp Facebook VK LinkedIn