Expert Tips: How to Deal with Nozzle Clogging in a Salt Spray Test Chamber

Source:LINPIN Time:2025-05-20 Category:Industry News

If your salt spray test chamber experiences spray failure, the issue may be a clogged nozzle. So, how should you handle nozzle clogging? Here’s a clever trick from our experts—a simple guide on how to resolve nozzle blockages in testing equipment.

As the saying goes, “The right remedy works best.” To quickly fix a clogged nozzle in a salt spray test chamber, you first need to identify the cause of the blockage. Possible reasons include:

Long-term disuse and lack of cleaning, leading to crystallization or chemical deposits that form blockages in the nozzle or its channels (a coupling effect where two or more substances combine).
Frequent use with a damaged filtration system, causing fluid to create an abrasive “sandblasting” effect that damages the nozzle.

salt spray test chamber
Physical impact, such as cracks or chips in the nozzle due to accidental collisions.
To diagnose the issue, check the pressure gauge. If the pressure is normal, the problem likely lies in the nozzle or tubing. Observe whether the saline solution flows properly to determine whether the blockage is in the nozzle or the pipeline.

If the nozzle is clogged with debris:

Use an air gun to blow out the blockage from the nozzle’s tip (never blow from the silicone tube side).
If there is no debris but poor airflow:

Turn on the spray switch and check if air flows through the curved glass tube of the nozzle. If not, adjust the nozzle’s position.
If the nozzle is damaged:

Replace it immediately to ensure proper operation.
Preventive Measures to Avoid Nozzle Clogging
To minimize nozzle blockages:

Use filtered or softened water (such as distilled or deionized water) in the test chamber to reduce crystallization.
Avoid overly small nozzle apertures, as they are more prone to clogging.
Install an oil-water separator to prevent moisture and oil from the compressed air from contaminating the nozzle.
Routine maintenance and careful handling can prevent accidental collisions that may damage the nozzle.
By following these steps, you can effectively prevent and resolve nozzle clogging, ensuring smooth and accurate salt spray testing. Stay tuned for more maintenance tips on testing equipment!

News Recommendation
Recently, LINPIN Instruments has successfully won the bid of Baowu Iron and Steel Company Limited's ultra-low temperature reliability environment testing project. This is another cooperation between the company and Baowu Iron & Steel following the cooperation of Anshan Iron & Steel, Hesteel, JISCO, Shagang and other iron and steel companies, which not only reflects the deep trust and tacit understanding established between the two sides in the long-term cooperation, but also lays down a colorful stroke for the diversified cooperation between the company and the iron and steel group.
Providing a stable and continuous voltage to the constant temperature and humidity test chamber is essential for ensuring its proper operation. To achieve a stable current supply, the device's power must be correctly connected to the circuit. So, how is this done?
To ensure relatively accurate test results from a rain test chamber, there are several key points to pay attention to. Below, we outline these precautions for your reference.
Thermal shock test chambers, renowned for their outstanding stability and reliability, are extensively employed in the testing of automation components, telecommunications equipment, plastics, aerospace hardware, chemical materials, LEDs, and electronic devices.
In today's rapidly developing technological era, many products need to maintain stable performance under different temperature conditions. The walk-in temperature and humidity test chamber has thus emerged.
Product Recommendation
Telegram WhatsApp Facebook VK LinkedIn