Introduction to the Location and Quantity of Test Points in Low-Temperature Test Chambers

Source:LINPIN Time:2025-04-02 Category:Industry News

A low-temperature test chamber can accommodate multiple test specimens, but are there specific requirements regarding the location and quantity of test points? Today, we will introduce the guidelines for test point placement and quantity in such equipment.

1. Three Standard Test Levels
Inside a low-temperature test chamber, there are typically three horizontal test planes: upper, middle, and lower.

The upper plane is usually positioned at 1/10 of the chamber’s height from the top.
The middle plane is located at the geometric center of the workspace.
The lower plane is placed 10 mm above the bottom sample rack.

low-temperature test chamber
2. Placement of Test Specimens
The primary test specimen (for central monitoring) should be placed at the geometric center of the workspace.
Other test specimens should maintain a distance of at least 1/10 of the chamber’s perimeter from the walls.
For chambers with a working volume ≤ 1 m³, the minimum distance should be ≥ 50 mm.
3. Relationship Between Chamber Capacity and Number of Specimens
If the chamber’s working volume is < 2 m³, the recommended number of test specimens is 9.
If the volume is ≥ 2 m³, the recommended number increases to 15.
The total number of specimens generally increases with chamber capacity to ensure representative testing.
Now that you have a basic understanding of test point placement and quantity in low-temperature test chambers, you may consult our technical experts for further details.

About Our Low-Temperature Test Chambers
Our low-temperature test chambers are designed to evaluate the storage and operational adaptability of electronic and other products under cyclic low-temperature conditions. These tests help optimize material selection, ensuring better performance in cold environments.

To meet growing market demand, we offer various chamber sizes with the following features:
✅ High-density insulation for superior thermal retention
✅ Microcomputer-controlled temperature adjustment with digital display
✅ High automation for precise and stable testing

We welcome all interested clients to visit our facility for equipment demonstrations and project discussions. Let’s collaborate to develop reliable testing solutions!

News Recommendation
There are many important considerations when selecting and using a formaldehyde/VOC environmental test chamber. How many of these are you aware of? Today, we’ll share a few key points—have you noticed them?
It is crucial to operate a high and low temperature test chamber correctly.
In generic reliability standards such as GB/T 2423, IEC 60068 and MIL-STD-810, “damp heat” is treated as an independent climatic stress. The goal is not merely to verify moisture resistance, but to accelerate and expose failure modes triggered by water adsorption, condensation, “breathing” and electrochemical migration.
Recently, China issued the Action Program for Promoting Large-scale Equipment Renewal and Consumer Goods Trade-in (hereinafter referred to as the Action Program), which systematically deploys 20 specific measures around the implementation of the four major actions of equipment renewal, consumer goods trade-in, recycling, and standard enhancement. Since the release of the Action Program, provinces and municipalities have responded to the policy by gradually promoting the relevant work and formulating provincial and municipal plans to promote large-scale equipment renewal and consumer goods trade-in.
The sealing of the low-temperature test chamber is crucial. If the sealing of the equipment is compromised, air leakage may affect the test results, so timely action is required. Here’s what you can do
Product Recommendation
Telegram WhatsApp Facebook VK LinkedIn