Vacuum drying oven, vacuum oven is specially designed for drying heat-sensitive, easy to decompose and easy to oxidize substances, which can be filled with inert gas to the interior, especially some items with complex composition can also be dried quickly. Widely used in electronics and chemical industry.

Service Hotline

4000-662-888
Product Features
Product Parameter
Model Studio Size  (D*W*H)mm Dimension(D*W*H)mm power
DZF-6020 300×300×275 605×490×450 450W
DZF-6050 415×370×345 710×560×550 1450W
DZF-6090 450×450×450 610×660×1400 1400W
DZF-6210 560×600×640 720×820×1750 2200W
temperature range RT+10℃~200℃
Temperature resolution ±0.1℃
Thermostatic fluctuation ±1℃
Vacuum level achieved 133Pa
timing range 1-9999M
Outer case material High quality A3 steel plate electrostatic spraying
Inner box material Imported high-grade stainless steel SUS304
Insulation material High-density fiberglass wool
viewing window Bulletproof Double Tempered Glass Door
temperature sensor PT100 platinum resistance temperature measurement body
standard configuration Sample rack 1 layer / 2 layers / 3 layers
security protection Over-temperature protection, test space temperature fuse, fan motor overheating, equipment timing
Leakage protection, the overall equipment under phase / reverse phase (only 380V power supply)
Supply Voltage AC220V±10% 50Hz 或AC380V±10% 50Hz
operating environment 5℃~+28℃ ≤85% RH
Note:
1、This technical information, subject to change without notice
2、Can be customized according to the specific requirements of the user non-standard vacuum drying oven etc

 

Product Recommendation
Industry News
A high and low-temperature test chamber is designed to simulate natural climatic conditions to test the high- and low-temperature resistance of products or materials under alternating temperature changes.
A low-temperature test chamber can accommodate multiple test specimens, but are there specific requirements regarding the location and quantity of test points? Today, we will introduce the guidelines for test point placement and quantity in such equipment.
Calibrating the temperature of a constant temperature and humidity test chamber is a preparatory task before testing. So, how can you ensure proper calibration of the test equipment? Only after adjusting the temperature of the chamber can testing proceed normally.
Telegram WhatsApp Facebook VK LinkedIn