How to Set the Timer Controller of a Sand and Dust Test Chamber

Source:LINPIN Time:2025-05-06 Category:Industry News

How are the parameters of the sand and dust test chamber timer controller set? Here’s a guide: Today, we’ll share the detailed steps for operating the timer controller.

Before operating the timer controller of the sand and dust test chamber, ensure that the power is connected, as these operations must be performed with the power on. Initially, the controller display will show the main menu: Run, Function Block, Clear Program, etc.
sand and dust test chamber

Access the Function Block

Press the Up/Down Arrow keys to move the cursor to Function Block, then press OK (Confirm).
The screen will display “T1” (Test Time) in the lower right corner, with the cursor blinking on “T1”.
Set the Test Time (T1)

Press the Left/Right Arrow keys to move the blinking cursor to the time setting position.
Observe the time unit displayed in the upper left corner of the screen.
Press SEL, then use the Up/Down Arrow keys to adjust the desired time value.
Press OK to confirm, and the cursor will return to “T1”.
Set Blowing Dust Time (T2)

Press the Up Arrow and SEL keys simultaneously—the screen will display “T2” (Blowing Dust Time, recommended: 5 minutes).
Follow the same method as T1 to set the time.
Set Blowing Dust Stop Time (T3)

Press the Up Arrow—the screen will display “T3” (Stop Blowing Dust Time, recommended: 2 minutes).
Set it using the same method as T1.
Set Vibration and Power Times (T4-T7)

T4: Vibration Time (recommended: 0.5 seconds).
T5: Stop Vibration Time (recommended: 2 minutes).
T6: Power-On Time (set according to relevant standards).
T7: Power-Off Time (also set according to standards).
Complete Setup and Start the Test

After setting all 7 time parameters, press ESC (Exit) to return to the main menu.
Use the Up Arrow to move the cursor to “Run”, then press OK—”Run” will change to “Stop”, and the internal power will activate.
The temperature controller will display readings. Press ESC again to return to the control screen.
Begin the Test

Turn on the “Blowing Dust” and “Vibration” switches—the device will start working.
To begin the test, turn on the “Sample” switch.
That’s all for the operation steps of the sand and dust test chamber timer controller. By reading this guide, you should now have a clear understanding of the process.

News Recommendation
Ozone aging test chambers are indispensable instruments in the field of material anti-aging testing. Common types include ozone aging test equipment and ultraviolet aging test equipment.
When a high and low temperature test chamber experiences refrigerant leakage, the following steps can be taken to address the issue
During the operation of a temperature and humidity test chamber, some abnormal situations may occur, such as excessive overheating of the condenser tube.
Enterprises should pay attention to both the needs of users, but also to pay attention to competitors, and strive to find a balance between user needs and competitors marketing concept, known as “market-oriented”, the core of which is from the user's needs Adhere to the market orientation, the pulse of the user, the foothold and the destination point on the basis of the product “sell”, based on the concept, LF Instruments positive development of 20 ℃/min fast temperature change test chamber and successfully introduced to the market. Based on this concept, LF Instruments has developed a 20℃/min fast temperature change tester and successfully introduced it to the market.
High and low temperature alternating test chambers actually come in various structural designs. No device is perfect, but each has its own set of advantages and disadvantages. Today, I will review the pros and cons of four different structural designs of these test chambers.
Product Recommendation
Telegram WhatsApp Facebook VK LinkedIn