How to Set the Timer Controller of a Sand and Dust Test Chamber

Source:LINPIN Time:2025-05-06 Category:Industry News

How are the parameters of the sand and dust test chamber timer controller set? Here’s a guide: Today, we’ll share the detailed steps for operating the timer controller.

Before operating the timer controller of the sand and dust test chamber, ensure that the power is connected, as these operations must be performed with the power on. Initially, the controller display will show the main menu: Run, Function Block, Clear Program, etc.
sand and dust test chamber

Access the Function Block

Press the Up/Down Arrow keys to move the cursor to Function Block, then press OK (Confirm).
The screen will display “T1” (Test Time) in the lower right corner, with the cursor blinking on “T1”.
Set the Test Time (T1)

Press the Left/Right Arrow keys to move the blinking cursor to the time setting position.
Observe the time unit displayed in the upper left corner of the screen.
Press SEL, then use the Up/Down Arrow keys to adjust the desired time value.
Press OK to confirm, and the cursor will return to “T1”.
Set Blowing Dust Time (T2)

Press the Up Arrow and SEL keys simultaneously—the screen will display “T2” (Blowing Dust Time, recommended: 5 minutes).
Follow the same method as T1 to set the time.
Set Blowing Dust Stop Time (T3)

Press the Up Arrow—the screen will display “T3” (Stop Blowing Dust Time, recommended: 2 minutes).
Set it using the same method as T1.
Set Vibration and Power Times (T4-T7)

T4: Vibration Time (recommended: 0.5 seconds).
T5: Stop Vibration Time (recommended: 2 minutes).
T6: Power-On Time (set according to relevant standards).
T7: Power-Off Time (also set according to standards).
Complete Setup and Start the Test

After setting all 7 time parameters, press ESC (Exit) to return to the main menu.
Use the Up Arrow to move the cursor to “Run”, then press OK—”Run” will change to “Stop”, and the internal power will activate.
The temperature controller will display readings. Press ESC again to return to the control screen.
Begin the Test

Turn on the “Blowing Dust” and “Vibration” switches—the device will start working.
To begin the test, turn on the “Sample” switch.
That’s all for the operation steps of the sand and dust test chamber timer controller. By reading this guide, you should now have a clear understanding of the process.

News Recommendation
Previously, we introduced some common problems with high and low temperature test chambers. Today, we will discuss the issue of slow cooling in the test equipment, which may be caused by malfunctions in the refrigeration system.
The safety of low-temperature test chambers during operation is of utmost importance. Not only must environmental safety measures be in place, but proper operational protocols must also be followed.
Low temperature test chambers serve as critical equipment for material low-temperature performance verification, electronic component reliability testing, and biological sample preservation.
The refrigeration compressor is one of the core components of a high/low-temperature test chamber, and its selection is closely related to the test results. In fact, semi-hermetic compressors are commonly used in these chambers. Why is this the case? Today, we will explore the three main reasons.
When maintaining cold and hot shock test chambers, paying attention to details is crucial. A minor oversight can significantly impact the effectiveness of the tests. Therefore, it is essential to prioritize the regular maintenance of testing equipment. Today, let's go over a few key aspects of maintaining test chambers.
Product Recommendation
Telegram WhatsApp Facebook VK LinkedIn