What to Do When the Circulating Water Pressure in a Thermal Shock Test Chamber is Insufficient?

Source:LINPIN Time:2025-04-14 Category:Industry News

If you find that the circulating water pressure in your thermal shock test chamber is insufficient, how should you resolve it? To address the issue at its root and prevent recurrence, we need to promptly identify and troubleshoot the problem based on the actual situation.

Step-by-Step Troubleshooting Guide:
Check the Circulating Water Indicator Light on the Display Panel

If the indicator light is flashing, it typically means there was a previous issue with the cooling water tower (e.g., power failure, malfunction in the water replenishment system, or the water inlet system not being activated). However, if the issue has already been resolved but the light remains flashing, further inspection is needed.

Thermal Shock Test Chamber
If the Indicator Light is Solid Red (Not Flashing)

This suggests a current problem requiring immediate troubleshooting. Possible causes include:
Cooling water tower malfunction
Water pump failure
Power supply interruption
Inspection Steps:
Check if the cooling water tower’s power is off → If so, turn it back on.
Verify if there is water in the cooling tower → If empty, ensure the water inlet valve is open and the system is functioning properly.
Examine the water pump (motor) → If the pump does not operate even when powered, it may be damaged and require repair or replacement.
Temporary Fix vs. Professional Assistance
The above methods allow operators to troubleshoot basic issues (though not as precisely as professionals).
If the problem persists after these steps, it’s best to contact the manufacturer’s technical support for further diagnosis and repair.
Our Support Services
As a decades-long manufacturer of testing equipment, we provide:
✔ Various test chamber models
✔ Technical support & solutions
✔ Expert troubleshooting assistance

For technical inquiries, call us at 400-066-2888. Our engineers will help you resolve the issue efficiently.

News Recommendation
The ozone aging test chamber, also known as the ozone aging test machine, is used for rubber products such as vulcanized rubber, thermoplastic rubber, cable insulation sheaths, etc.
If the temperature recovery time falls outside the specified range, it can affect the test results of the thermal shock chamber. To routinely test the temperature recovery time of the equipment, you can follow the steps below.
How are the parameters of the sand and dust test chamber timer controller set? Here’s a guide: Today, we’ll share the detailed steps for operating the timer controller.
Dust testing, codified in GB/T 2423.37 and IEC 60068-2-68, is now mandatory for proving the sealing integrity and reliability of military, automotive, photovoltaic and rail-transit products. Direct test outlays—consumables, energy, labour and depreciation—seldom exceed one to three per cent of the selling price, yet hidden expenses triggered by poor chamber selection—repeat tests, invalid data or after-sales downtime—can leap to eight to twelve per cent.
Product reliability is no longer governed solely by design rules and manufacturing precision; it is equally decided by a product’s ability to survive sudden climatic extremes. Among all environmental stressors, rapid temperature change is the single most frequent root-cause of field failures in electronic components, automotive assemblies, and aerospace hardware.
Product Recommendation
Telegram WhatsApp Facebook VK LinkedIn