Expert Tips: How to Deal with Nozzle Clogging in a Salt Spray Test Chamber

Source:LINPIN Time:2025-05-20 Category:Industry News

If your salt spray test chamber experiences spray failure, the issue may be a clogged nozzle. So, how should you handle nozzle clogging? Here’s a clever trick from our experts—a simple guide on how to resolve nozzle blockages in testing equipment.

As the saying goes, “The right remedy works best.” To quickly fix a clogged nozzle in a salt spray test chamber, you first need to identify the cause of the blockage. Possible reasons include:

Long-term disuse and lack of cleaning, leading to crystallization or chemical deposits that form blockages in the nozzle or its channels (a coupling effect where two or more substances combine).
Frequent use with a damaged filtration system, causing fluid to create an abrasive “sandblasting” effect that damages the nozzle.

salt spray test chamber
Physical impact, such as cracks or chips in the nozzle due to accidental collisions.
To diagnose the issue, check the pressure gauge. If the pressure is normal, the problem likely lies in the nozzle or tubing. Observe whether the saline solution flows properly to determine whether the blockage is in the nozzle or the pipeline.

If the nozzle is clogged with debris:

Use an air gun to blow out the blockage from the nozzle’s tip (never blow from the silicone tube side).
If there is no debris but poor airflow:

Turn on the spray switch and check if air flows through the curved glass tube of the nozzle. If not, adjust the nozzle’s position.
If the nozzle is damaged:

Replace it immediately to ensure proper operation.
Preventive Measures to Avoid Nozzle Clogging
To minimize nozzle blockages:

Use filtered or softened water (such as distilled or deionized water) in the test chamber to reduce crystallization.
Avoid overly small nozzle apertures, as they are more prone to clogging.
Install an oil-water separator to prevent moisture and oil from the compressed air from contaminating the nozzle.
Routine maintenance and careful handling can prevent accidental collisions that may damage the nozzle.
By following these steps, you can effectively prevent and resolve nozzle clogging, ensuring smooth and accurate salt spray testing. Stay tuned for more maintenance tips on testing equipment!

News Recommendation
As is well known, the inner chamber of a UV aging test chamber is typically made of stainless steel, while the outer chamber is constructed from steel plates coated with paint. The choice of stainless steel material can significantly impact the test results, while the selection of paint affects the device's appearance and usability.
High and low temperature test chambers have a lifespan from the moment they are produced.
To ensure the accuracy of test results from a low-temperature test chamber, the following points should be noted during operation
The performance of the compressor in a high-low temperature alternating test chamber directly affects the efficiency of the equipment and the test results. Therefore, customers should choose a high-quality compressor to enhance the device's performance. So, how can you determine whether the compressor in the test equipment is good or bad?
Some companies use salt spray test chambers very frequently, and long-term use inevitably leads to malfunctions.
Product Recommendation
Telegram WhatsApp Facebook VK LinkedIn