Troubleshooting Temperature Cycling Test Chambers: Are You Using the Right Methods?

Source:LINPIN Time:2025-04-16 Category:Industry News

When a temperature cycling test chamber malfunctions, how should you troubleshoot it? Have you identified the correct diagnostic approach? Mastering the right methods can help quickly locate the issue and restore functionality more efficiently.

Common Faults and Solutions:
Chamber Does Not Start When Powered On:

Check if the main circuit breaker is engaged or damaged. Replace it if necessary.
Overheating Issue:

Possible cause: Solid-state relay is short-circuited.
Solution: Replace the relay and check if the control parameters are malfunctioning. Adjust the settings if needed.

temperature cycling test chamber
Slow Heating:

Possible causes:
Fan not working properly → Inspect the fan.
Chamber door not closed tightly → Ensure proper sealing.
Excessive load → Reduce the load.
No Heating at All:

Possible causes:
Solid-state relay is open-circuit → Replace the relay.
Heating element is damaged → Replace the heating tube.
No Cooling:

Possible causes:
Solid-state relay is short-circuited → Replace the relay.
Insufficient refrigerant → Replenish the refrigerant.
Temperature Fluctuation (Unstable Control):

Possible causes:
PID parameters are misconfigured → Recalibrate the PID settings.
Temperature sensor failure → Replace the sensor.
Overpressure Alarm:

Possible causes:
High ambient temperature → Lower the surrounding temperature or replace the axial fan.
Excess refrigerant → Release some refrigerant.
Solution: Press the reset button and check the system.
Overload Protection Triggered:

Possible causes:
Faulty overload protector → Replace it.
Excessive current → Adjust the overload range and press the thermal overload reset button.
Also, inspect the refrigeration circuit for issues.
These are some of the key troubleshooting methods for temperature cycling test chambers. If the problem persists after applying these solutions, contact after-sales support for professional assistance.

News Recommendation
Cooling is a critical aspect of constant temperature and humidity test chambers. The two most common cooling methods are water-cooled (chilled water) and air-cooled. Each has its own advantages and disadvantages.
Conducting tests on specimens using sand and dust test chamber equipment incurs certain costs. One of the most critical factors influencing the cost of specimen testing is the selection of the testing equipment. How does the specific equipment affect operating costs?
A salt spray test chamber is an instrument used to evaluate the corrosion resistance of objects to salt spray. After testing, how can users determine whether the samples meet the required standards? The judgment criteria vary depending on the applicable standards. Once the relevant standard is identified, the following methods can be used for evaluation.
Cold and hot shock test chambers utilize high-temperature and low-temperature zones during testing. In the testing process, two-box models involve placing the test product in high and low-temperature zones for cyclic exposure, while three-box models involve blowing high- and low-temperature gases onto the product alternately.
Providing a stable and continuous voltage to the constant temperature and humidity test chamber is essential for ensuring its proper operation. To achieve a stable current supply, the device's power must be correctly connected to the circuit. So, how is this done?
Product Recommendation
Telegram WhatsApp Facebook VK LinkedIn