How to Address Humidity Issues in High and Low Temperature Alternating Humidity Test Chambers?

Source:LINPIN Time:2025-05-09 Category:Industry News

As a device designed to simulate temperature and humidity variations for quality and adaptability testing of products or materials, the high and low temperature alternating humidity test chamber can also be used for material stress screening and evaluating various resistances, such as heat resistance, moisture resistance, and cold resistance. This testing equipment features a long service life, high safety, environmental friendliness, energy efficiency, and strong reliability, meeting consumer testing needs in R&D, product acceptance, and quality inspection.

high and low temperature alternating humidity test chamber

Common Humidity Issues and Solutions
During humidity testing, if the humidification system fails to operate, the actual humidity may deviate significantly from the target humidity, while the temperature difference remains minimal. If the humidity fails to reach the desired level, the following aspects should be checked:

Dry Wet-Bulb Sensor Gauze

Insufficient humidity may be caused by a lack of moisture on the wet-bulb sensor gauze.
Check if the gauze is properly placed and saturated with water.
Low Water Level in the Water Tank

The water level in the tank is automatically controlled by a water level controller.
If the tank is low on water, the humidity will not reach the target value.
Inspect the water supply system to ensure it is functioning correctly.
Faulty Water Level Controller

If the water level controller malfunctions or the supply system fails to deliver water, humidity control will be affected.
Aged or Hardened Wet-Bulb Gauze

If the gauze has been used for too long, it may lose its ability to absorb water effectively.
Replace the gauze with a new one to restore proper humidity control.
⚠ Note:

The test chamber requires high-purity water (distilled or deionized).
Using tap water can cause the gauze to harden over time, reducing its water absorption capacity.
Troubleshooting Steps
Check the wet-bulb gauze – Replace if hardened or dry.
Inspect the water tank – Ensure sufficient water supply.
Verify the water level controller – Confirm proper operation.
Test the water supply system – Check for blockages or malfunctions.
If the issue persists after these checks, contact a professional testing service provider. Shanghai Linpin offers comprehensive technical support to ensure your test chamber operates optimally.

For further assistance, feel free to consult our technical team! 🚀

News Recommendation
Improper operation of a formaldehyde test chamber may lead to equipment failure. To help users avoid operational errors, here are some important points to keep in mind when using the device.
Although both thermal-shock and steady-state temperature/humidity tests fall under the umbrella of “weathering verification”, they differ fundamentally in physical modelling, control logic and safety strategy.
After prolonged use of a temperature and humidity test chamber, certain components or systems may develop issues, such as sediment buildup in the water system. To ensure accurate test results, regular cleaning of the water system is essential.
When a high and low temperature test chamber experiences refrigerant leakage, the following steps can be taken to address the issue
Intelligent control is crucial for rain test chambers.
Product Recommendation
Telegram WhatsApp Facebook VK LinkedIn