How to Resolve Unstable Low Temperatures in High and Low Temperature Alternating Test Chambers

Source:LINPIN Time:2025-05-29 Category:Industry News

Regularly, staff should master some troubleshooting and handling methods for high and low temperature alternating test chambers. If the equipment exhibits unstable low temperatures, what causes this phenomenon, and how should we address these faults?

Check the Electrical System of the Equipment
If the refrigeration compressor can start normally during operation, it indicates that the issue is not related to the electrical system.

high and low temperature alternating test chambers

Verify Whether the Refrigerant in the Refrigeration System Is Sufficient
During operation, inspect the temperature of the compressor’s exhaust and suction pipes. If the exhaust pipe temperature is not high and the suction pipe temperature is not low, it suggests a lack of refrigerant.

Inspect the Main and Auxiliary Units
When both the main and auxiliary units are operating simultaneously, the cooling rate of the equipment is relatively high. In the initial stage of temperature maintenance, both units work together. Once the temperature stabilizes, the main unit alone maintains the cooling and stability, while the auxiliary unit stops. If the main unit has a refrigerant leak, its cooling efficiency will weaken. Since both units operate together during cooling, unstable temperatures will not occur, but the cooling rate will decrease. During temperature maintenance, if the main unit fails to cool and the auxiliary unit is inactive, the temperature inside the chamber will gradually rise. When it reaches a certain level, the control system will activate the auxiliary unit to lower the temperature. Once the temperature drops sufficiently, it re-enters the maintenance phase. This cycle repeats, resulting in unstable low temperatures.

After identifying the cause of unstable low temperatures in the test chamber, what measures should be taken to resolve it?

For electrical system issues, technicians should repair the system.
If the refrigerant is insufficient, technicians should replenish it.
For problems with the main and auxiliary units, a combination of soapy water and leak detection devices can be used to locate leaks. Once found, the leaks should be repaired or the components replaced.

News Recommendation
Previously, we have discussed many technical aspects of temperature cycling test chambers. Today, we will share methods for detecting leaks in these chambers. Specifically, there are four main approaches to check for leaks.
The performance indicators of a high and low temperature test chamber are crucial because they directly affect the reliability of test results, the breadth of application scenarios, and the effectiveness of quality assurance for products.
In order to verify the ability of satellite components to withstand vacuum thermal environments, a thermal vacuum chamber is used for testing. The purpose is to validate the rationality of the design and identify defects in the components, materials, and manufacturing processes.
Frost formation in a thermal shock test chamber is a sublimation phenomenon where moisture inside the chamber condenses under low-temperature conditions. But why does this happen? Is it simply because the chamber temperature is too low?
To meet various market demands, many types of testing chambers have been developed, among which the salt spray test chamber is one of the most common.
Product Recommendation
Telegram WhatsApp Facebook VK LinkedIn