Why Does the Compressor of a Low-Temperature Test Chamber Fail to Start?

Source:LINPIN Time:2025-03-24 Category:Industry News

The compressor is a critical component of a low-temperature test chamber. Why does it sometimes fail to start? What aspects should we check?

The troubleshooting process for a low-temperature test chamber compressor that fails to start should begin with an electrical circuit inspection. First, check if there is an issue with the motor circuit connection. If the problem lies in the equipment’s circuit, inspect the motor circuit connection points to determine if poor connections are the cause. Immediate fault diagnosis and repair are necessary.
low-temperature test chamber
Additionally, check if the power supply voltage is too low. If the grid voltage is insufficient, wait for it to return to normal levels before attempting to start the compressor.

Another factor to consider is whether the exhaust valve is leaking. If the exhaust valve is damaged or improperly sealed, it can cause excessive pressure in the crankcase, preventing the compressor from starting. In this case, the exhaust valve and sealing components need to be replaced.

Malfunctions in the energy regulation mechanism can also cause the compressor to fail to start. This requires checking the oil supply pipeline for blockages, low pressure, or stuck oil pistons. If any faults are detected, they should be addressed promptly.

Pressure relays and temperature controllers should also be inspected. If the issue lies with the pressure relay, the set pressure parameters may need adjustment. If the temperature controller is malfunctioning, it should be recalibrated or replaced if damaged. In such cases, it is advisable to contact the manufacturer’s technical support for repair or replacement.

Low-temperature test chambers are widely used in industries such as aerospace, military, construction materials, scientific research, and electronics. As a result, many clients in these industries purchase them. Linpin Instruments is a manufacturer of low-temperature test chambers and can design and produce equipment based on customer requirements. The equipment is provided directly by the manufacturer, offering more competitive prices. The manufacturer also provides delivery, technical support, and other services to ensure comprehensive customer satisfaction. For inquiries, please call 4000-662-888 to learn more about the equipment.

News Recommendation
The rain test chamber is a testing instrument used to evaluate the effectiveness of water resistance.
The refrigeration compressor is one of the core components of a high/low-temperature test chamber, and its selection is closely related to the test results. In fact, semi-hermetic compressors are commonly used in these chambers. Why is this the case? Today, we will explore the three main reasons.
To ensure the accuracy of test results from a low-temperature test chamber, the following points should be noted during operation
The refrigeration system is a crucial component of a high and low-temperature humidity test chamber, and its normal operation is vital for the accuracy of test results. Here are some simple methods to help you determine whether the refrigeration system is working properly。
Cooling is a critical aspect of constant temperature and humidity test chambers. The two most common cooling methods are water-cooled (chilled water) and air-cooled. Each has its own advantages and disadvantages.
Product Recommendation
Telegram WhatsApp Facebook VK LinkedIn